Description
- With solid membrane Interchangeable flat surface
- Ion to be detected passes through the membrane generating a difference of potential, proportional to ionic concentration
- Analysis time is generally less than one minute
- Sample preparation steps such as filtering or Distillation rarely necessary
- Calibration should be carried out as often as possible (once a day) at the same temperature as the sample
- Measurement temperature 0 to 50°C
- Delivered with membrane, electrolytic solution, ISAB solution and ISE standard and gasket
ISAB solution:
To adjust the ionic strength. add directly to the solution to be measured.
ISE standard:
For the calibration of ionic electrodes. Concentration 1000 ppm.
Specifications
Specifications
| Applications | Determination of NO3- ion |
| Filling Solution Type | ISAB/ISE |
| pH Range | 2.5 to 11 |
| Interferences | CIO4-, I-, CN-, BF4-, Cl-, NO2- |
| Electrode | Nitrate Combined ISE |
| Parameter | Nitrate Ion |