missing translation for 'onlineSavingsMsg'
Learn More
Learn More
Description
- Ready-to-use column for convenient process development optimization and parameter screening
- Highly cross-linked agarose matrix with average bead size of 90μm
- Bed height of 10cm designed to allow method optimization and parameter screening
- 20cm bed height is acheived when it is connected in series
- Small bed volume (4.7mL) gives fast result and minimal sample/buffer consumption
- Reproducible results, scalable to BioProcess columns packed with the same chromatography media using the same linear fluid velocity
- Bed dimension: 7.7 × 100mm
- Matrix: 6% cross-linked agarose beads
- Average particle size: 90μm
- Max. flow velocity: 450cm/hr.
- Material (column hardware and column tube): polypropylene
- Column I.D.: 7.7mm
HiScreen Q FF
- Media:
- pH stability working range: 2 to 12
- pH stability cleaning: 2 to 14
- Ion exchanger type: strong anion exchanger
- Media:
- pH stability working range: 3 to 14
- pH stability cleaning: 4 to 13
- Ion exchanger type: strong cation exchanger
Specifications
Specifications
| Buffer | 1M NaOH, 8M Urea, 8M Guanidine Hydrochloride, 70% Ethanol |
| Product Type | HiScreen SP Sepharose FF clumn |
| Bed Volume | 4.7mL |
| Diameter (English) Inner | 0.3 in. |
| Diameter (Metric) Inner | 7.7mm |
| Quantity | 1 x 4.7mL |
| Height (English) | 3.93 in. |
| Height (Metric) | 100mm |
| Flow Rate | <450 cm/hr. |
| Material | Polypropylene |
| Show More |
Safety and Handling
Product Identifier
- HiScreen SP FF
- Warning
- Flammable liquid Category 3
- H226-Flammable liquid and vapour.
- P210-Keep away from heat, hot surfaces, sparks, open flames and other ignition sources. No smoking.
- P241-Use explosion-proof electrical/ventilating/lighting/ equipment.
- P260-Do not breathe dust/fume/gas/mist/vapours/spray.
- P280-Wear protective gloves/protective clothing/eye protection/face protection.
- P305+P351+P338-IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present and easy to do. Continue rinsing.
- P501b-Dispose of contents/container in accordance with local/regional/national/international regulations.
- MIXTURE LIST-Contains : ethanol
- UNKNOWN ACUTE-18 per cent of the mixture consists of component(s) of unknown acute toxicity.